Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
145 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
1.4. CEPH sensor: Bright calibration
1. Acquire Sensor Bright Calibration in Norm and Fast Mode respectively.
2. Run the C:\VCaptureSW\Acquisition\CEPH\WideCEPH\Norm\VACAP.exe file.
3. The standard of Lv value for each bright calibration point is as follows.
Norm Mode Bright standard value: 5 points with Cu 1 mm filter
Bright Min Max
Median
value
Cal Point 1 200 300
Cal Point 5 1200 ~ 1400
Fast Mode Bright standard value: 5 points with Cu 1 mm filter
Bright Min Max
Median
value
Cal Point 1 200 300
Cal Point 5 600 ~ 800