3. Acquisition Module Error Code
Error code sending method
Method: Broadcast Message
MsgName: K-DEVICE
WParam : HIWord = Main Code (If less than 0, an error)
LoWord : 0xAAAA
LParam : HIWord = Sub Code
LoWord : 0xAAAA or Tail Code
[Code : E003.000.000]
Multilingual error phrases output
Main
Code
Sub
Code
Tail
Code
Description
E003 ------ ------ Acquisition Module related error
000 ------ Success
010 ------ [ Initialization failure related ]
000 Can’t open the COM port.
001 Can’t initialize frame grabber interface or can’t reserve memory for
acquisition.
002 Can’t communicate to MCU or Modem Ring signal has an improper
state.
020 ------ [ Capture failure related ]
000 Capture error.
030 ------ [ Reconfiguration failure related ]
000 Can’t initialize COBRA COM-server.
001 Reconstruction error. Some bugs in VXM-file or insufficient memory.
061 ------ [ HW Error related ]
HW
Error
No
[smp_errr value transfer]: HW Error is transmitted when an equipment
error occurs while the Acquisition module is running. (refer to the HW
Error Code)

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals