PaX-i3D Green (PHT-60CFO) User Manual
156
Appendices
3 Electromagnetic Compatibility
(EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions.
The PaX-i3D Green (Model: PHT-60CFO) is intended for use in the
electromagnetic environment specied below. The customer or the user of
the PaX-i3D Green (Model: PHT-60CFO) should assure that it is used in
such an environment.
Immunity Test Compliance
Electromagnetic Environment -
Guidance
RF Emissions
CISPR 11
Group 1
The PaX-i3D Green (Model: PHT-
60CFO) uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
(The PaX-i3D
Green (Model:
PHT-60CFO)
in combination
with the shield
location)
The PaX-i3D Green (Model: PHT-
60CFO) is suitable for use in all
establishments other than domestic,
and may be used in domestic
establishments and those directly
connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes, provided the following
warning is heeded:
Warning: This equipment/
system is intended for use by
healthcare professionals only. This
equipment/ system may cause radio
interference or may disrupt the
operation of nearby equipment. It
may be necessary to take mitigation
measures, such as re-orienting
or relocating the PaX-i3D Green
(Model: PHT-60CFO) or shielding
the location.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
uctuations/
Flicker
emissions
IEC 61000-3-3
Complies
NOTE) It is essential that the actual RF shielding effectiveness and filter
attenuation of the shielded location be veried to ensure that they meet or
exceed the specied minimum values.