PaX-i3D Green (PHT-60CFO) User Manual
80
6 Acquiring PANO Images
When you click the
CONFIRM
button,
The
READY
button will begin blinking to show that it has been activated.
The rotating unit will move to its initial scanning position.
Three laser beams,
Mid-sagittal plane, Frankfurt plane and Canine
, will
be activated to facilitate patient positioning. These beams will disappear
after 20 minutes or if the
READY
button is clicked.
NOTE
Laser beam On / Off buttons:
: Imaging Program
: Control Panel
The scan time and DAP (Dose Area Product) values will be shown on the main
display for the selected exposure.
J. Guide the patient to the equipment. Position the patient within the
equipment. For further information about patient positioning, refer to section
6.3
Positioning the Patient.
[PI3DG_130U_44A_en]User Guide.indd 80 2016-05-24 오후 4:18:56