3 PaX-i3D Green Imaging System Overview
3.3.3 Emergency Stop Switch
During operation, the following emergency situations may occur:
X-ray emission continues after the exposure switch has been released
Injury to the patient or damage to the equipment
Other emergency situations
If a problem occurs during image acquisition, press the red
Emergency Stop
Switch
to immediately stop the moving parts and cut off all power to the
equipment’s electrical components. To reset the
Emergency Stop Switch
, turn
it clockwise until it pops up.
3.3.4 The Exposure Switch
The exposure switch allows the operator
to control image acquisition from outside
of the X-ray room. Press and hold the
exposure switch down until acquisition
is complete. Premature release of
the exposure switch will abort image
acquisition.
Pressing the exposure switch activates the
orange indicator light to indicate that the
X-ray is being emitted.
Exposure Button
Exposure Indicator light
Green: Ready
Orange: X-ray On
[PI3DG_130U_44A_en]User Guide.indd 39 2016-05-24 오후 4:18:33