Vitalograph Spirotrac Instructions for Use
IFU Part No.: 07402
Issue 22
_______________________________________________________________________________________________________________
Page 33 of 78 DT_0006-15
On receiving access to the Vitalograph Fusion Quality Assured Spirometry
System the user will be provided with the over-read functionality.
3.17.1 Setup mail settings
Spirotrac must be setup with a suitable mail server in order to send data to an over-
reader.
1. Select Tools > Options from the Main Menu. Select Mail Setup.
2. Enter the details required as follows:
a. SMTP Server The mail server name with your Internet Service Provider
(ISP)
b. SMTP Port The mail server port number with your Internet Service
Provider (ISP), typically but not always: port 25.
c. Connection The dialup connection to establish during transmission, if
required.
d. SMTP Username The user ID of your account with the ISP, if a login is
required.
e. SMTP Password The user password of your account with the ISP, if a
login is required.
f. Over-Read E-Mail Address The email address of the over-reader to
which any session data is to be sent.
3. The connections button will take the user to the System Dial-up connection
manager.
4. A test button is provided to check settings.
3.18 Reference Curve and Data
Spirotrac allows the user to compare the best test (curve and data) from a user
defined reference session. When set, the reference test will appear in both the data
and graph results.
1. To enable Reference Curve functionality select Tools > Options > General.
2. Click Yes for the Reference Curve and Data option. (To disable the
functionality, click No.)
3. To set a session as a reference session right click on the required session and
select the Set as Reference option.
3.19 Health Level 7 Interface
Health Level 7 (HL7) is an internationally recognised standard for the exchange and
management of electronic healthcare information between Electronic Medical Record
(EMR) systems and Spirotrac.
More information about using the HL7 interface can be obtained from technical support
at +353 65 6864082 or email spirotrac@vitalograph.ie
3.19.1 Changing Health Level 7 (HL7) Settings within Spirotrac
1. Select Tools > Options > Advanced.
2. Check HL7 as enabled to enable HL7. (Uncheck to disable.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vitalograph Spirotrac 7000 and is the answer not in the manual?

Vitalograph Spirotrac 7000 Specifications

General IconGeneral
BrandVitalograph
ModelSpirotrac 7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals