4-4
IM WT5000-02EN
Setting Independent Integration (Independent Control)
4.
Tap Independent Control. Element numbers appear.
5.
Tap within the element number display frame. The following screen appears.
Computation/Output tab
Set the integration conditions.
Select the input elements that you want
independent integration to be performed on.*
Selects all input elements
Clears all input elements
Even if you select input elements that independent integration will be performed on, independent
integration may not be performed due to the wiring system setting or the independent input element
configuration (see the features guide).
*
Element number display
Setting the Watt-Hour Integration Method for Each Polarity (WP±
Type)
4.
Tap WP± Type. The following screen appears.
Select the integration method setup method (Each, All).
When you select Each, you can set the integration method for each input element.
Set the integration method.
(Charge/Discharge, Sold/Bought)
4.1 Setting Integration Conditions

Table of Contents

Other manuals for YOKOGAWA WT5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT5000 and is the answer not in the manual?

YOKOGAWA WT5000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT5000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals