KRYPTON®
TECHNICAL REFERENCE MANUAL
8.2.3. SAFE-OP state 
In this state the EtherCAT master is already reading the data from the slice/s at a slow sampling rate: the
D LED is slowly blinking. DeweSoft will set the Krypton slice/s in the SAFE-OP state when you are in
Channel Setup. 
SAFE-OP state
8.2.4. OP state
In this state the EtherCAT master is reading the data from the slice/s at the desired sampling rate: the D
LED is turned ON. DeweSoft will set the Krypton slice/s in the OP state when you are in Measure Mode.
OP state
KRYPTON® V20-1 110 / 141