6173.3 Babylog 8000/8000 plus/8000 SC 12/00 General Page 12
For internal use only. Copyright reserved.
GBK61733XXTL1.fm 07.07.05
11 Leak Test
Adjust the following parameters on the Babylog 8000/
Babylog 8000 plus/Babylog 8000 SC (see Figure below):
Fig. 6: Leak Test
•Select pressure from the menu “pressure curve” and “measured values”.
Switch off occurring alarms and warnings by pressing the “Reset/Check” key.
Press and hold the manual inspiration key.
A pressure is now built up for 5 s. During this time, a pressure of at least 75 mbar
must be achieved.