Dräger Medical AG & Co. KGaA 6173.3
Function description Babylog 8000
7 Cold Start/Warm Start Behavior
A distinction is made between an intentional switch-off with the power switch and a power failure. If a
power failure is recognized, the position of the low voltage contacts of the power switch is evaluated.
7.1 Cold-Start Behavior
When Babylog is switched off, all set parameters and data on the power switch state are stored. When
switching on the Babylog, this information is reloaded and a "cold start" is carried out. Babylog initializes,
carries out its function tests, indicates the results of these tests, and starts ventilation.
7.2 Warm-Start Behavior
A "warm start" is carried out after a power failure. Babylog does not carry out a self-test, but starts
ventilation immediately. It continues to work with the previously set parameters. A visible and audible
alarm is active during power failure.

Table of Contents

Other manuals for Dräger Babylog 8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Babylog 8000 and is the answer not in the manual?

Dräger Babylog 8000 Specifications

General IconGeneral
Leak CompensationYes
TypeVentilator
Pressure Range0 to 80 mbar
Tidal Volume Range2-2000 mL
Frequency1-150 breaths/min
Inspiratory Time0.1-3.0 sec
PEEP0 to 20 mbar
FiO2 Range21-100%
Power Supply100 to 240 V AC, 50/60 Hz

Related product manuals