Function description Babylog 8000
9.5 Patient System Heater
The patient system heater consists of a heating resistor located at the expiratory-valve connection.
9.6 Pressure Sensor Base PCB
The Pressure Sensor Base PCB has four pressure sensors for the measurement of
Pinsp
, P
exp
, AIR and
O
2
supply pressures and for coordination of the pressure and the O
2
measurement signals.
9.7 O
2
Amplifier PCB
The O
2
Amplifier PCB amplifies the sensor signals directly at the sensor.