28
Dräger Medical AG & Co. KGaA 6173.3
Function description Babylog 8000
6.2 Keys
6.2.1 CPAP
Pressing this key will activate the CPAP mode. The current respiratory cycle is stopped before a new
mode becomes active (for example, IPPV/IMV). This key is protected against unintentional operation by a
software-controlled locking. After power-on, Babylog will automatically enter the most recent mode
selected.
6.2.2 IPPV/IMV (CMV) (up to software version 4.n)
Pressing this key will activate the controlled ventilation mode. This key is protected against unintentional
operation by a software-controlled locking. If CPAP was set before, the first mandatory breath is applied
after the last expiratory phase has been completed (zero expiratory flow) or after a zero flow has been
measured at the Y-piece for one second.
The CPAP and IPPV/IMV keys deactivate each other. The selected ventilation mode is indicated by a
green LED on the respective key.
6.2.3 Man. Insp.
Pressing this key will apply a breath at the set inspiratory flow rate and the set pressure limit (Pinsp). This
breath is stopped when the key is released or when a fixed time limit of 5 s is reached. The next manual or
mandatory breath can only be applied after an expiratory phase (IPPV/IMV) has been completed. If a
manual breath is stopped by the time limit, the next mandatory or manual breath can only be applied after
a fixed time-out (5 s).
6.2.4 2-min Silence
Pressing this key will silence the audible alarm for 2 minutes. This status is indicated by the yellow LED
on the key. Pressing the key again within the silence time will reactivate the audible alarm.
6.2.5 Reset/Check (OK)
Pressing the Reset/Check key (or OK key as of software version 5.n) will confirm or cancel text messages
shown on the display. The message currently shown on the display will be cancelled and the audible
alarm will be silenced.
Pressing the key for a longer period of time (approx. 2 s) will start a display test and an audible alarm
generator test.
6.2.6 Backlight On/Off (Babylog 8000/8000 SC with LC Display)
Pressing this key will switch the LCD backlight on or off.

Table of Contents

Other manuals for Dräger Babylog 8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Babylog 8000 and is the answer not in the manual?

Dräger Babylog 8000 Specifications

General IconGeneral
Leak CompensationYes
TypeVentilator
Pressure Range0 to 80 mbar
Tidal Volume Range2-2000 mL
Frequency1-150 breaths/min
Inspiratory Time0.1-3.0 sec
PEEP0 to 20 mbar
FiO2 Range21-100%
Power Supply100 to 240 V AC, 50/60 Hz