CADENCE II Fetal Monitor User Manual
- 23 -
4.7 Alarm Defaults
Alarm Setting Options Default
FHR1/FHR2/DECG Alert
On, Off On
FHR1/FHR2/DECG Lower Limit
50 bpm ~ 130 bpm, in increments of 5. 120 bpm
FHR1/FHR2/DECG Upper Limit
150 bpm ~ 210 bpm, in increments of 5. 160 bpm
FHR1/FHR2/DECG Keep Time
0 ~ 30 second(s), in increments of 5. 30 seconds
US1/US2/DECG Probe Detecting
On, Off On
FHR1/FHR2/DECG Signal Quality
Detecting
On, Off On
UA Probe Detecting
(Toco transducer detecting)
On, Off On
SOV (signals overlap verification)
On, Off Off