CADENCE II Fetal Monitor User Manual
- 54 -
2) Follow the manufacturer’s instruction to dilute the solution, or adopt the lowest
possible density.
3) Do not immerse any part of the monitor or any accessory into liquid.
4) After disinfection, no remaining disinfectant is allowed on the surface.
5) Check if the monitor and accessories are in good condition. If any aging or damage
is detected, replace the damage part(s) or contact the manufacturer for service
before reusing them.
NOTE:
The manufacturer has no responsibility for the effectiveness of controlling infectious
disease using these chemical agents. Please contact infectious disease experts in your
hospital for details.
8.4 Sterilizing
Do not sterilize the monitor or the accessories, unless this is necessary according to your hospital
regulation.