EasyManuals Logo
Home>EDAN>Monitor>CADENCE II

EDAN CADENCE II User Manual

EDAN CADENCE II
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
CADENCE II Fetal Monitor User Manual
- 60 -
The Maximum Charge
Current Continuously
2000mA
Storage
Short Term (within 1 month): -20 ºC to 60 ºC (-4 ºF to 140 ºF)
Middle Term (within 3 months): -20 ºC to 45 ºC (-4 ºF to 113 ºF)
Long Term (within 1 year): -20 ºC to 20 ºC (-4 ºF to 68 ºF)
During storage, recharge the battery at least every six months.
Charge State When
Leaving Factory
20% ~ 50%
Cycle Life 500 times
A1.6 Low Output Summary Table
Low Output Summary Table
(for systems with no transducers having global maximum index values exceeding 1.0)
System: CADENCE series Fetal Monitor
Transducer Model
I
spta.3
(mW/cm
2
)
TI Type TI Value MI
Ipa.3@MI
max
(W/cm
2
)
TIS 0.055
PW 1.0 MHz 2.77
TIB 0.629
0.029 0.01