CADENCE II Fetal Monitor User Manual
- 45 -
1) Insert the remote event marker connector into the MARK socket on the monitor.
2) Let the patient hold the marker in hand; ask her to press the top key of it when a fetal
movement is felt. Continuous movements in 5 seconds are considered to be one
movement and only press the key once.
7.7.7 Clearing FM Count
When you want to clear the FM count, you can either press the AUTO key to start a new
monitoring, or
1 Press the SETUP key once to open the Function menu.
2 Press the UP key or DOWN key to move the cursor to ClearFM.
3 Press the CHANNEL key to confirm selection.
4 Press the AUTO key to exit.
7.8 Login
7.8.1Inputting Patient’s Bed ID
The bed ID you input for the patient will be printed on the recorder paper.
1 Press the SETUP key three times to open the Fetal Monitor Configuration menu.
2 Press the CHANNEL key when the cursor stays at BedID.
3 Press the UP key or DOWN key to choose the bed ID from 1 ~ 99.
4 Press the CHANNEL key to confirm selection.
5 Press the AUTO key to exit.
7.8.2 Inputting Patient’s Information
It’s recommended to input the patient’s information. This information will appear beneath the
loaded trends.
1 Press the SETUP key once to open the Function menu.
2 Press the CHANNEL key when the cursor stays at Login.
3 An ID that is composed of time or the previous ID appears in the ID area. Press the
CHANNEL key, the cursor moves to the symbols area below. Press the UP key or
DOWN key to move the cursor left or right. Press the FREEZE key to move the cursor
up and down. Press the CHANNEL key at the
symbol to delete an inputted symbol.
Press the CHANNEL key at the
symbol to finish inputting. You can input 10
characters at most.
4 Press the UP key or DOWN key to move the cursor to NAME, press the CHANNEL key.
Input the patient’s name in the same way. You can input 10 characters at most.