Emerson Process Management GmbH & Co. OHG 5-1
X-STREAM Non-Incendive
Instruction Manual
HASXENE-IM-EX
12/2014
5
Dismounting
Chapter 5
Dismounting and Disposal
5.1 Dismounting and Diposal of the Analyzer
HAZARDS FROM DISMOUNTING
Dismounting instruments installed in hazardous area requires special
instructions to be followed!
Only trained personnel, observing all applicable technical and legal
requirements, and aware of the possible risks is permitted to dismount
these analyzers.
Failure to follow may result in explosion, death or personal injury!
EXPLOSION HAZARD WHEN OPEN
Dismounting requires to open the instrument and is permitted only if no
hazardous atmosphere is present and both the instrument and connected
circuitry are de-energized!
Depending on the local regulation this may require a competent hot work
supervisor to issue a hot work permit.
Failure to follow may result in explosion!
HEAVY INSTRUMENT
X-STREAM fi eld housings, intended for outside and wall mounted use, weigh
between 26 kg (57 lb) and 45 kg (99 lb) depending on options installed.
Support the instrument before loosening the fi xing screws (risk of dropping
instrument).
Two people and/or lifting equipment is required to lift and carry these units.
Failure to follow may result in personal injury!