Emerson Process Management GmbH & Co. OHG2-2
X-STREAM Non-Incendive
Instruction Manual
HASXENE-IM-EX
12/2014
2.2 General Safety Instructions
2.2 General Safety Instructions
EXPLOSION HAZARD
Consider all applicable standards and legislative requirements when
installing the analyzer, and connecting power and signal cables. Take
care of the relevant installation standards, as there are (but not limited to)
e.g. EN 60079-14 (Europe), National Electrical Code (NEC-NFPA 70; USA),
Canadian Electrical Code (CEC; Canada), IEC 60079-14 (International), and
all corresponding standards.
Consider all instructions given in this, and all associated analyzer manuals!
Read and understand all instructions given in the manuals before beginning
to install the analyzer!
Failure to follow the instructions may cause explosion, property damage
and personal injury or death!
EXPLOSION HAZARD
Installing this instrument requires opening the housing and working at open
instrument. This is permitted only if no hazardous atmosphere is present,
and the instrument and all connected external ciruitry is de-energized!
X-STREAM gas analyzers do not provide a power switch and are operable
when connected to power!
Depending on the local legislation, this may require a hot work permit to
be issued.
EXPLOSION HAZARD BY GASES
X-STREAM analyzers utilize not only sample gas but one or more pressurized
carrier gases and/or calibration gases.
If external equipment (e.g. a fl owmeter for fl ow control) is required, legislative
requirements and instructions for installation in hazardous (classifi ed) areas
must be considered.