Emerson Process Management GmbH & Co. OHG4-10
X-STREAM Non-Incendive
Instruction Manual
HASXENE-IM-EX
12/2014
Gases
Zero gas: 0.000 ppm
Span gas: 50000.000 ppm
Range gases..
Current range: Range 1
4.6.1 Preparing Calibrations
Before performing calibrations it is essential
to tell the instrument the calibration gas con-
centrations.
Starting from the MEASUREMENT SCREEN
press
DOWN to open the MAIN MENU, enter
SETUP-CALIBRATION and directly enter
GASES.
Enter the concentration values for zero and
span gases to be used for calibration:
• Values may be found on gas supplier‘s
certifi cates.
• Proper calibrations require proper gas
concentration entries!
• Multi-channel instruments require to enter
separate values for each channel.
Multi-channel unit:
Select the channel to be calibrated in
SELECT COMPONENT.
Setup..
Calibration..
Gases..
Ch2
Tolerances
DeviatToler: Off
Zero Limit: 20.0 %
Span Limit: 20.0 %
Press LEFT to return to CALIBRATION, and
then enter TOLERANCES.
By default, „DeviatToler“ is disabled (Off).
With „DeviatToler“ enabled (On):
• while calibrating, the analyzer compares
the set values for zero or span gas with
the actual value.
• calibration is cancelled if set value and
actual value differ by more than the values
in lines 2 and 3, thus preventing miscali-
brations by mistakenly wrong calibration
gas supply (e.g. if span gas instead of zero
gas is connected for zero calibration).
4.6 Perform a Calibration