Maintenance
4.6 Perform a Calibration
OPERATION AT LOW TEMPERATURES
When operating an instrument at temperatures below 0 °C (32 °F), do
NOT apply gas nor operate the internal pump before the warmup time has
elapsed!
Violation may result in condensation inside the gas paths or damaged
pump diaphragm!
Before performing any actions, make sure the required calibration gas is applied
and fl owing!
Supply all calibration gases with the same fl ow and pressure as the sample gas (re-
commended: approx. 1 l/min), and utilizing
the correct gas fi tting.
Ensure the warm-up time after switching on has elapsed! Warm-up time depends on
installed measuring system and confi guration, measurement specifi cations in
section 1.7!
Do NOT calibrate the TRACE MOISTURE sensor (tH
2
O)!
The sensor is completely calibrated with all calibration data stored in its fl ash memory
and does not require recalibration:
If the sensor is included into a calibration procedure, it might end up with a wrong ca-
libration and unusable sensor. Therefore the analyzer's trace moisture measurement
channel is confi gured to be excluded from autocalibration procedures, by default calibra-
ting all channels. This exclusion is done by factory setup and cannot be changed.
For proper measurement results we recommend to exchange the sensor regularly
after 12 months of operation. For instructions on how to exchange,
X-STREAM
Enhanced instruction manual
Do NOT calibrate the TRACE OXYGEN sensor (tO
2
) without prior reading the
instructions!
Together with each sensor an installation manual is shipped, also giving comprehensive
calibration information.
Read these information prior intending to activate calibration procedures!