Emerson Process Management GmbH & Co. OHG 4-13
X-STREAM Non-Incendive
Instruction Manual
HASXENE-IM-EX
12/2014
4
Maintenance
Zero Calibration..
Span Calibration..
Advanced Calibration..
Apply Gas..
Lock Menus!
Acknowledgements
Pump 1: Off
Pump 2: Off
Ranges..
4.6 Perform a Calibration
When nished, press
LEFT several times to re-
turn to SELECT COMPONENT (multi channel
analyzer only), to perform a span calibration
for another channel,
or
press
HOME to return to the MEASUREMENT
SCREEN, to nish with manual calibration
procedures.
4.6.3 Manual Span Calibration
Multi-channel unit:
Select the channel to be calibrated in
SELECT COMPONENT.
Before selecting any further line
make sure the required calibra-
tion gas is applied and owing!
Span calibration offers the same options as
zero calibration.
Select "Span Calibration.."
Span Calibration
Cancel!
Start!
Span Gas 0.000 ppm
Concentration 0.000 ppm
Flow 0.00 l/min
Status..
Results..
Restore..

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM Series and is the answer not in the manual?

Emerson X-STREAM Series Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals