1.3 Technical Data
1.3 Technical Data
1.3.1 Installation Site and Protection Method
Test samples have been approved by independent test institutes according to the relevant stan-
dards and found to comply with the requirements.
The certifi cates can be viewed in the appendix of this manual (
page A-1).
Admissible Values
Humidity
(non-condensing)
< 90 % RH at +20 °C (68 °F)
< 70 % RH at +40 °C (104 °F)
Degree of pollution 2
Installation category II
Elevation 0 to 2000 m (6560 ft) above sea level
Ambient atmosphere Non corrosive
Protection Concept: Ex nA nC
North America (CSA-c/us)
Hazardous area Zone 2, Gas Div 2
Classifi cation Class I, Zone 2, Gas,
AEx nAC IIC T4
Ex nAC IIC T4
Class I, Div 2,
Group ABCD
Temperature class T4 T4
Underlying stan-
dards
I CAN/CSA-
E60079-0: 02 (R2006)
CAN/CSA-
E60079-15:02 (R2006)
ANSI/ISA-12.00.01
-2002 (IEC 60079-0
Mod)
ANSI/ISA-12.22.01-
2002 (IEC 60079-1
Mod)
UL 60079-15:2009
CSA C22.2 No 213-
M1987
ISA 12.12.01-2007