2.6 Installation - Gas Connections
2.6 Gas Connections
EXPLOSION HAZARD BY FLAMMABLE GASES
Take care that all external gas lines are connected as described and that
they are gastight to avoid leakages! Also, gas lines for fl ammable gases
must end in a safe area!
Faulty connected gas lines lead to explosion hazard or even to mortal
danger!
HAZARD BY HARMFUL TO HEALTH GASES
Take care that all external gas lines are connected as described and that
they are gastight to avoid leakages! Faulty connected gas lines lead to
mortal danger!
Don‘t breathe the emissions! Emissions may contain harmful to health or
toxic components, and may cause headache, sickness, unconsciousness
and death.
Dispose emissions by a safe gas fl ue and check its function periodically!
Do not confuse gas inlets and outlets. All gases supplied must be prepared
beforehand. When supplying aggressive gases, ensure that the gas lines
are not damaged.
Max. admissable pressure: 150 kPa!
Exhaust lines must be installed to incline downwards and be unpressurized
and protected against frost, and conform to legal requirements.
More information is provided in the related X-STREAM series instruction
manual.