Startup
3.9 Checking the Settings
*) Numbers 1 to 5 refer to components [chan-
nels] 1 to 5: In case of secondary parameters,
this means, the selected value is that of the
sensor assigned to the given component
(Press2 is the pressure value of the sensor
assigned to component 2).
In contrast, capital letters A to D imply that
these calculator results are component
[channel] independent (Calc C is the result
of calculator C).
Signal
*)
Description
None
The analog output is deactivated
0 mA
A 0 or 4 mA signal is generated, e.g. to check the signal processing in an
external system. Whether a 0 or 4 mA signal is generated, is set by the "Out
range” line (
next page).
20 mA
A 20 mA signal is generated, e.g. to check the signal processing in an exter-
nal system.
Comp1 ... 5 Gas concentration
Temp1 ... 5 Temperature
Press1 ... 5 Pressure
Flow1 ... 5 Flow
CalcA ... D Result of calculator
RawVal1 ... 5 Raw value
RangeID1 ... 5 ID of selected range

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM Series and is the answer not in the manual?

Emerson X-STREAM Series Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals