130 AK 98™ Dialysis Machine - Maintenance support
HCEN12630 Revision 07.2015
Program version 1.xx
5.3.3.3 To edit preset files offline
With the Gambro software AK 98 Preset Editor installed on a PC you can edit preset
files offline.
Install AK 98 Preset Editor on the PC. Export the preset files from the AK 98 dialysis
machine to a USB flash drive and then import them to the PC.
Refer to Section 5.3.3.1 “To export presets” on page129, Section 5.3.3.2 “To import
presets” on page129 and the AK 98 Preset Editor - User Guide for more information.
5.3.3.4 To configure multiple dialysis machines
Perform the following to configure multiple dialysis machines with the same presets:
Procedure
1) Preset one AK 98 dialysis machine with the correct presets.
2) Export the settings to a USB flash drive. Refer to Section 5.3.3.1 “To export
presets” on page129.
3) Import the settings from the USB flash drive to the other machines, one at the
time. Refer to Section 5.3.3.2 “To import presets” on page129.
5.3.4 Default preset
The Default preset should be treated with care, as it will erase all customised presets
and restore factory default values.
Procedure
1) Enter Preset mode, see Section 5.1.3 “Enter service or preset mode” on
page123.
2) Select the Default tab.
3) Press Set Default.
4) Press Confirm.
5.4 Service mode
5.4.1 Service mode overview
In the Service mode you can access the:
Calibration tab to calibrate the transducers and variable flow, set the clock, and
calibrate the touch screen. Refer to Section 5.4.2.1 “Calibration menu” on
page131.
Utility tab to clear the chemical disinfection status or disinfection/decalcification
attentions, perform service tests or drain, export log data to USB flash drive or
import log config from a USB flash drive. Refer to Section 5.4.3.1 “Utility tasks” on
page131.
Diagnose tab to scan for hardware updates on the i2c bus, view and erase
technical errors, perform tests and calibrations, and enter settings. Refer to
Section 5.4.4.1 “About diagnose” on page133.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals