262 AK 98™ Dialysis Machine - Functional check
HCEN12630 Revision 07.2015
Program version 1.xx
Conditions Evaluations & Actions Error codes
Open TAVA. N/A N/A
Measure the PD pressure. Is the PD between -220 to
-180 mmHg? If not - error
code.
0511 088 067
Table 12-24. Valves leakage test - phase 9
12.7.3.2 Fluid path - Leakage test
This test shall be performed after exchanging the CHVA and CBVA valves. All
concentrate connectors shall be connected to the machine. BiCart cartridge holder
shall be closed.
Conditions Evaluations & Actions Error codes
INVA, HBVA, ZEVA,DIVA,
BYVA, TAVA, EVVA, AIVA,
REVA Closed
N/A N/A
RIVA, FLVA Open N/A N/A
Flow pump stopped N/A N/A
CHVA open and CBVA closed Suction pump regulating HPG
to -350 mmHg.
If not achieved within 90
seconds then error code 0511
088 115
Closing RIVA. Suction pump is
set to constant speed creating
a negative pressure of approx.
-550 mmHg outside the latch
pressure. Wait 4 seconds to
stabilize pressure.
HPG reference value stored. If HPG ref value > -250 mmHg
then error code 0511 088 116
Table 12-25. Fluid path - Leakage test

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals