HCEN12630 Revision 07.2015
Program version 1.xx
AK 98™ Dialysis Machine - Calibrations 163
Use following formula to calculate which gauge pin to use:
2 x wall thickness x 0,7 (mm)
7.3.5.2 Adjustment instruction
Check the range between the blood pump rollers and the path according to Figure 7-7
“Blood pump” on page163. Adjust range between the blood pump rollers and the
path if it is necessary.
Figure 7-7. Blood pump
The GO pin should barely pass within the calibrating area.
The STOP pin is NOT allowed to pass in any point within the calibrating area.
NOTE!
Put grease on the GO and STOP pins when not used and store gauge in plastic bag
to prevent it from corroding.
7.3.6 Blood leakage detector calibration
NOTE!
In order to maintain stable temperature and to prevent disturbances caused by light
admission, do not remove the BM during the calibration of the blood leakage
detector.
Procedure
1) Make sure there is no fluid in the AK 98 dialysis machine monitor. If there is fluid
in the monitor, press the Disinfection button to drain the machine.
2) Remove the blood leakage detector cover and clean the inside of the blood
leakage detector housing.
3) Mount a blood leakage calibration cover (K40420002). Make sure the filter is in
horizontal position (the knob on the outside of the blood leakage calibration
cover).

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals