HCEN12630 Revision 07.2015
Program version 1.xx
AK 98™ Dialysis Machine - Functional check 273
Conditions Evaluation & actions Error code
UF Supervision taration time is
measured.
UF Supervision time duration
from UF Cell taration
completion shall not exceed
10min.
0511 088 131
12.7.11 Blood leak clean test
This test is performed by enabling the LED. The value should be <30. If not an
attention will be issued:
507 Blood leak detector failure
Clean detector and press Confirm.
12.8 Automatic restart of the FM functional check
In some cases when an error code is issued, it can be enough to restart the machine
to make the monitor pass the functional check. To avoid the need of manually restart
an automatic restart of the FM functional check is performed.
Instead of generating a technical error, the monitor automatically restarts the fluid part
of the functional check. A silent attention is generated informing the operator that the
functional check has been restarted. The attention is active until green fluid path is
achieved or confirmed by the operator. As a consequence of the restart of the FM
functional check, the functional check is prolonged.
In addition a silent error code (no technical alarm) is stored in the error code buffer
indicating which test caused restart of the functional check. If the same error occurs
again in the following functional check, the monitor will issue an error code the usual
way.
The restart of FM functional check is used for the following error codes in functional
check:
Temperature tests DIVA/TAVA TEST Unstable degassing
pressure
0511 088 055 0511 088 061 0505 006 003
0511 088 056 0511 088 065 N/A
0511 088 092 0511 088 086 N/A
N/A 0511 088 088 N/A
Table 12-33. Error codes in functional check
Silent error codes stored in the error buffer due to a hydraulic restart:
Temperature tests DIVA/TAVA TEST Unstable degassing
pressure
0512 088 055 0512 088 061 0512 006 003
0512 088 056 0512 088 065 N/A
0512 088 092 0512 088 086 N/A
N/A 0512 088 088 N/A
Table 12-34. Silent error codes

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals