270 AK 98™ Dialysis Machine - Functional check
HCEN12630 Revision 07.2015
Program version 1.xx
Figure 12-18. AIVA test conditions
1. Deaerating chamber
2. Air inlet
3. Dropping fluid level
High level active - 0505 018 004
General description of conditions for occurring:
AIVA and REVA opened, INVA and HBVA closed.
Suction pump active and fluid tubes connected to safety bypass.
Low level in deaerating chamber is not achieved within 90 seconds.
The status from level sensor indicates high level in chamber.
High level active - 0505 018 004
General description of conditions for occurring:
AIVA and REVA opened, INVA and HBVA closed.
Suction pump active and fluid tubes connected to safety bypass.
Low level in deaerating chamber is not achieved within 90 seconds.
The status from level sensor indicates high level in chamber.
High level and low level not active - 0505 018 005
General description of conditions for occurring:
AIVA and REVA opened.
Suction pump active and fluid tubes connected to safety bypass.
Low level in deaerating chamber is not achieved within 90 seconds.
The status from level sensor indicates middle level in chamber.
12.7.9 DIVA/TAVA/EVVA closing test
Procedure
1) EVVA/DIVA is opened, TAVA/BYVA closed.
2) Control is requested to build up a negative pressure of -200 mmHg.
3) Protective waits at most 10 seconds to get a high level bubble in deairating
chamber.
4) Protective waits at most 10 seconds for PD_PRESSURE to reach below 0 mmHg.

Table of Contents

Other manuals for Gambro AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gambro AK 98 and is the answer not in the manual?

Gambro AK 98 Specifications

General IconGeneral
BrandGambro
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals