2.2.9 Step 9 - Ultrafilter (option)
WARNING!
Check carefully that there is no leakage from the ultrafilter after changing it.
WARNING!
Before a new treatment can be performed after changing the ultrafilter, the machine
must go through a disinfection to ensure the quality of the dialysis fluid.
NOTE!
The ultrafilter (UFD) is not included at delivery of the AK 98 dialysis machine.
NOTE!
Make sure that the ultrafilter is handled in an aseptic way according to the
corresponding package insert.
Procedure
1) Pull the handle and press the latch downwards to release the lower part of the
ultrafilter holder.
2) Insert the ultrafilter into the holder and push it gently upwards.
3) Push the lower latch into position to close the holder.
4) Put a label on the ultrafilter with the installation date.
Figure 2-7. Ultrafilter (option)
2.2.10 Step 10 - Dialyzer holder
Procedure
1)
Install the dialyzer holder according to Figure 2-8 “Dialyzer holder” on page26.
2) Install and tighten the knob.