GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
5-16 Section 5-3 - Block Diagrams and Theory
5-3-3 General Information
LOGIQ™ S6 is a phased and linear array ultrasound imaging scanner. It has provisions for analog
input sources like ECG and phono. A dedicated Doppler probe may also be connected and used.
The system can be used for:
- 2D Black and White imaging
- 2D Color Flow
- M-Mode Black and White imaging
- Color M-Mode
- Doppler
- a number of combinations of the above
LOGIQ™ S6 is a digital beam forming system that can handle up to 192 element linear probes
through multiplexing.
Signal flow from the Probe Connector Panel to the Front End, to the Mid Processors and Back End
Processor and finally to the monitor and peripherals.
System configuration is stored on a hard disk and all necessary software is loaded from the hard
disk on power up.