DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 4-2 - General Procedure 4-19
4-3-6 Cineloop Check (QG)
For activating Cine, creating and storing Cine Loops, and information on the Cine Timeline, refer to the
LOGIQ™ S6 Quick Guide.
4-3-7 Backend Processor Checks
If all the previous tests have been passed successfully, the backend processor is most likely OK.
If the system seems to be operating erratically, please refer to Chapter 7, Diagnostic/
Troubleshooting.
4-3-8 Probe/Connectors Usage (QG)
The LOGIQ™ S6 Quick Guide, provides information on connecting, activating, deactivating and
disconnecting probes.
4-3-9 Peripheral Checks
Check that peripherals work as described below:
4-3-9-1 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
Table 4-40 Peripheral checks
Step Task to do Expected Result(s)
1.
Press Freeze Stop image acquisition.
2.
Press (P1) or (P4) on the Control panel
The image displayed on the screen is
printed on B&W or Color printer depending
on the key assignment configuration.
Table 4-41 GE Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2 B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
5 InSite/iLinq
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
6 Camera
Verify hardcopy output of the film camera. Clean as necessary.
7 Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
8ECG
Verify basic operation with customer
9 3D Probe
Ask Engineering about Calibration Check for 3-D

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals