GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
2-8 Section 2-3 - Facility Needs
2-3-2 Required Features
Dedicated single branch power outlet of adequate amperage (see Table 2-10) meeting all local and
national codes which is located less than 2.5 m (8 ft.) from the unit’s proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
NOTE: The LOGIQ™ S6 has three outlets inside the unit for on board peripherals.
Power outlets for other medical equipment and gel warmer
Power outlets for test equipment and modem within 1 m (3.2 ft.) of unit
Clean and protected space to store transducers (in their cases or on a rack)
Material to safely clean probes (done with a plastic container, never metal)
2-3-3 Desirable Ultrasound Room Facilities
Door is at least 92 cm (3 ft.) wide
Circuit breaker for dedicated power outlet is easily accessible
Sink with hot and cold water
Receptacle for bio–hazardous waste, like used probe sheaths
Emergency oxygen supply
Storage for linens and equipment
Nearby waiting room, lavatory, and dressing room
Dual level lighting (bright and dim)
Lockable cabinet ordered by GE for its software and service manual

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals