GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 6-6 - Jumper and Dip Switch Setting 6-25
Section 6-6
Jumper and Dip Switch Setting
6-6-1 Dip Switch Setting
Dip switche setting represents the version and revision of the board.
This section describes how to read the dip switches in case that a service engineer might need to
confirm and communicate the setting of the dip switches.
Normally the dip switches shall not be adjusted because those are set properly when the system or the
board is shipped out from the factory. In case of special needs, such as scanner upgrade, refer to each
instructions/documentations for proper setting..
Figure 6-103 A Example of DipSwitch Setting
Figure 6-104 Bit Setting for DIP Switch 1 to 4
4 (lower 4 bits)
E (Upper 4 bits)
Example: “E4 (h)” is set as shown.
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
0 (h) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h)
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
O
N
1
2
3
4
5
6
7
8
8 (h) 9 (h) A (h) B (h) C (h) D (h) E (h) F (h)
Bit Setting for DIP Switch 1 to 4 (lower 4 bits)

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals