DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
6-24 Section 6-5 - Cleaning the Trackball
Section 6-5 Cleaning the Trackball (cont’d)
8.) Insert the trackball into the housing.
9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner
retainer and trackball from the keyboard.
10.)Install the trackball retainer ring onto the inner retainer, then rotate it clockwise until its notches are
set in the horizontal position.
Figure 6-101 Removing Inner Retainer and Trackball
Figure 6-102 Rotating the Retainer Ring
Inner Retainer
Trackball
Inner Retainer
Stopper
Retainer Ring
Notches (Set Horizontally)