EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ S6

GE LOGIQ S6 User Manual

GE LOGIQ S6
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
4-26 Section 4-6 - Power Supply
Section 4-6
Power Supply
4-6-1 Power Supply Test Procedure
There is no need to perform any special tests on the Power Supplies if there does not seem to be a
problem that may be related to the Power Supplies.
4-6-2 Power Supply Adjustment
There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a unit
connected to that specific power outlet.