DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
3-6 Section 3-1 - Overview
3-2-2 Moving into Position
In general, a single adult can move the LOGIQ™ S6 along an even surface with no steep grades. At
least two people should move the machine when large humps, grooves, or grades will be encountered.
(It is better to pull from the rear rather than push from the front of the unit). Before moving, store all loose
parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.
Although LOGIQ™ S6 is a compact and mobile machine, two people should move it over rough
surfaces or up and down grades.
CAUTION
Do not lift the unit by the Keyboard.
Do not tilt the unit more than 10 degrees to avoid tipping it over.
To avoid injury by tipping over. Set the monitor to the lowest position before moving.
CAUTION
Equipment Damage Possibility. Lifting the console by holding covers may damage the covers.
Do not lift the console by holding any covers.