GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
3-8 Section 3-3 - Preparing for Installation
Section 3-3
Preparing for Installation
3-3-1 Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-3-2 Physical Inspection
3-3-2-1 System Voltage Settings
• Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ™ S6
Scanner is found on the label onto the rear lower of the scanner.
3-3-3 EMI Protection
This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
WARNINGWARNING
CONNECTING A LOGIQ™ S6 SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY THE SCANNER.
WARNINGWARNING
PROTECTIVE EARTH MUST BE TAKEN WHEN CONNECTING AC POWER CABLE
(200V) WITHOUT ITS PLUG TO WALL OUTLET.