Chapter 2. Installation
4 Moisture Target Series 6 Hygrometer User’s Manual
2.2 Selecting the Recorder Output
Note: By default, the recorder is set to the current output.
Note: The customer will provide their own cable for connecting the
recorder. Acceptable cables range from 16 to 26AWG.
The MTS6 has one isolated analog recorder output. The recorder output
provides either a current or voltage signal, which is set by switch S1 on the
main PC board.
Complete these steps to check or reset switch
S1 (see Figure 8 on page 7).
WARNING! Never connect line voltage or any other power input to
the recorder output terminals.
1. Make sure the MTS6 is turned off and unplugged.
WARNING! The MTS must be isolated or disconnected from all voltage
sources before changing the recorder output.
2. Remove the screw at the top of the back panel (see Figure 4).
Figure 4: Back Panel