Chapter 3. Operation and Programming
Moisture Target Series 6 Hygrometer User’s Manual 65
3.6.6 Reading and Setting the Calibration References
Note: The following procedure applies only if an M Series probe is being
used. If a VeriDri probe is used, Cal Reference... is not accessible.
IMPORTANT: The MTS6 is factory programmed with high and low reference
MH values. These values are generated from a factory lab
calibration and should not be changed without first consulting
GE technical support. Changes to these values will alter the
accuracy of the unit measurements.
To update calibration data, from the Settings
Menu select Cal Data and press Enter .
The following screen appears.
Note: To access the Cal Data menu, the User
Passcode is required (see section 3.6.1 on
page 56).
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the passcode has been entered,
press Enter a screen similar to the
following appears.
To view and/or edit the calibration reference
settings, select Cal Reference and press
Enter . The following screen appears.