Chapter 3. Operation and Programming
66 Moisture Target Series 6 Hygrometer User’s Manual
3.6.6a Setting the Calibration High Reference
To update high reference settings, from the
Edit Cal Refs menu select High Reference
and press Enter . A screen similar to the
following appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the changes have been made,
press Enter . The screen returns to the
Edit Cal Refs menu.
3.6.6b Setting the Calibration Low Reference
To update low reference settings, from the
Edit Cal Refs menu select Low Reference
and press Enter . A screen similar to the
following appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Once the changes have been made,
press Enter . The screen returns to the
Edit Cal Refs menu.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Moisture Target Series 6 and is the answer not in the manual?

GE Moisture Target Series 6 Specifications

General IconGeneral
BrandGE
ModelMoisture Target Series 6
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals