Chapter 5. Specifications
88 Moisture Target Series 6 User’s Manual
5.2 Moisture Measurement (cont.)
Probe Cable Length:
M Series: 2,000 ft (600 m) maximum
VeriDri: 1,000 ft (300 m) maximum
Moisture Probe Pressure Rating:
M1: 5 microns Hg to 75 psig (5 barg)
M2: 5 microns Hg to 5,000 psig (345 barg)
VeriDri: 5 microns Hg to 5,000 psig (345 barg)
Dew/Frost Point Temperature:
Overall Calibration Range:
–110° to 60°C
Available Calibration Range Options:
Standard: –80° to 20°C with data to –110°C
Extended High: –80° to 60°C with data to –110°C
Accuracy:
±2°C from –65° to 60°C
±3°C from –110° to -66°C
Repeatability:
±0.5°C from –65° to 60°C
±1.0°C from –110° to -66°C