3-7-4-5 UP-D23MD Printer - Settings
1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-4 "Adjustment of Printer Settings".
2.) Select the
PAPER page and set settings as shown in the left image at Figure 3-45 on page 3-45.
3.) Select the
GRAPHICS page. From the “Color Adjust” pop-up menu select:
a.) Color Balance: Cyan = 0; Magenta = 0; Yellow = 0
b.) Gamma Select: Gamma 1
c.) Color Correction: set Printer Hardware Color Correction
d.) Lightness: Sharpness = 7 or 8; Dark = 0; Gamma = -12; Light = 8
4.) Save the adjusted printer settings with APPLY
and OK. Close window and exit System Setup.
5.) Assign the Printer to the remote keys P1
, P2, P3 and/or P4; see: Section 3-7-5 on page 3-48.