GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-23
Section 8-15
Replacement of the Caps for Hardkeys
NOTE: The FRU part depends on Console (RTU).
Console version (Rafi or Whanam) is shown in the System Info page (see: Figure 7-1 on page 7-3).
8-15-1 Replacement of Circle Key Caps only
NOTE: The FRU part depends on Console (RTU).
Console version (Rafi or Whanam) is shown in the System Info page (see: Figure 7-1 on page 7-3).
8-15-1-1 Manpower
One person, approx. 1 minute/cap
8-15-1-2 Tools
small-sized slotted screwdriver or tweezers
8-15-1-3 Circle Key Caps - Replacement Procedure
1.) By means of a small slotted screwdriver, carefully push against the circle hardkey cap.
2.) Lift the cap, until it is completely loosened from its base.
3.) Place the new hardkey cap down until it snaps into position.
NOTICE
!! NOTICE:
Replacement procedure depends on key caps that have to be replaced!
• if just the circle key caps have to be replaced, refer to: Section 8-15-1 on page 8-23
• if trackball buttons or mode key slices have to be replaced too, please contact your local distributor
or GE service representative.
Figure 8-25 push against the circle cap and lift it

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
BrandGE
ModelVoluson E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals