GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
3-52 Section 3-8 - System Configuration
3-8-1-6 How to change Video Norm
1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-49.
2.) On the left side of the screen select CONNECTIVITY
and then click on the PERIPHERALS tab.
3.) If not currently selected, click the desired field PAL (50Hz) or NTSC (60Hz).
4.) Click SAVE&EXIT
to save Settings and exit System Setup.
5.) The system will reboot.
3-8-1-7 How to change Recorder Type
1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-49.
2.) On the left side of the screen select CONNECTIVITY
and then click on the PERIPHERALS tab.
3.) Depending on the used Recorder type, click the field VCR, DVD (Sony, JVC or internal DVR-Drive).
4.) Click SAVE&EXIT
to save Settings and exit System Setup.
3-8-1-8 How to adjust function of the Footswitch
1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-49.
2.) On the left side of the screen select CONNECTIVITY
and then click on the PERIPHERALS tab.
3.) Select desired function of the Footswitch.
4.) Click SAVE&EXIT
to save Settings and exit System Setup.
3-8-1-9 How to change the Keyboard Layout
see: Chapter 6 - Modification of Keyboard Layout, on page 6-9.
3-8-1-10 How to configure Service Platform
see: Chapter 7 - How to configure the Service Platform, on page 7-15.
Figure 3-53 System Setup - Connectivity - PERIPHERALS page
select
Recorder Type
select
Video Norm
Foot
Switch