7-22 Section 7-8 - Troubleshooting Trees, Instructions and Tech Tips
7-8-1 System does not boot up
Figure 7-25 System does not Power On / Boot Up
System doesn’t boot
Is Standby Switch
illuminated ?
Verify that
- Power Cord is connected to Mains
- Power Switch is turned on
(on the backside of the system)
Is Standby Switch
orange ?
Press Standby Switch till it
changes to green
System starts booting
Press Standby Switch
for approximately 5 sec.
until system is down.
Check cables between
Console and Monitor
Is there any picture on
the monitor ?
Is there any picture on
the touch screen ?
Shut off the power switch
Wait for 10 sec.
Shut on the power switch again
Is there any picture on
the monitor ?
Other error. Please
contact the service
engineer

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Power Requirements100-240 V AC, 50/60 Hz
ApplicationsObstetrics, Gynecology, Cardiology
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler
Transducer Ports4 active transducer ports
Image StorageInternal hard drive, USB storage
Transducer TypesLinear, Endocavitary
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals