GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
7-30 Section 7-8 - Troubleshooting Trees, Instructions and Tech Tips
7-8-7 Network Troubleshooting
7-8-7-1 No Connection to the Network at All
1.) Check that the network cable between the Voluson E8 / Voluson E6 system and the wall network
is connected and well seated in both ends. (Use a network cable that is known to be OK.)
2.) Connect a network cable between the system and a PC by either using a hub or a cross-over cable.
Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK.
7-8-7-2 GE remote service connection
The customer gets visual information about GE remote service status (shown in the status area on the
right bottom of the screen).
Gray = Cable disconnected or no network signal on a connected cable.
Green = Cable connected to a network. Does not imply
proper network settings.
Gray = Idle State Remote access is inactive. No InSite services are running.
Green = Active State
Disruptive Mode and Virtual Console Observation (VCO) is enabled, but there are no
active incoming connections.
Red = Disrupted State
Remote Access is active. All processes [UL_VNC and UL_CSD] are active.
In this state the Voluson E8 / Voluson E6 system should NOT be used clinically.

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
BrandGE
ModelVoluson E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals