GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
7-32 Section 7-9 - Error Messages
Section 7-9
Error Messages
NOTICE
!! NOTICE:
If the problem (error message) still exists after performing the described actions, call technical support.
Error Messages Actions
3D B_RxLines exceeded! Action: restart the system restart the system
3D B-O_Frames exceeded! Action: restart the system restart the system
3D C_RxLines exceeded! Action: restart the system restart the system
3D C-O_Frames exceeded! Action: restart the system restart the system
530-Probe connected on Connect another Probe
530-Probe connected on left probe connector (A). Connect another Probe
530-Probe connected on middle probe connector (B). Connect another Probe
530-Probe connected on right probe connector (C). Connect another Probe
530-Probe connected on upper-left probe connector (CW). Connect another Probe
A USB-drive has been removed without stopping it. Unplugging or ejecting devices
without first stopping them can often cause crashes and lose valuable data.
press Ok, and stop it before removing it next time
AcquMode - incorrect BBC_Wnd pointer restart the system
Another drive is already assigned to "xx" that is no DVD/CD rec.
Cannot swap assignment.
Use a DVD/CD writer
Aperture Shape Form for 1.5D Array wasn't found !!! Action: restart the system restart the system
Application setting ApodTable field count mismatch restart the system
Application setting dynamic set field count mismatch restart the system
Application setting field count mismatch restart the system
Application setting RxFrqResponse field count mismatch restart the system
AVI Save function fails Action: check the connection and Power cable - restart the
machine and try again
see error message
B_Enhance Out Of Range press ok and save this user-setting once again
B_Gain Out Of Range press ok and save this user-setting once again
B_Reject Out Of Range press ok and save this user-setting once again
B_TxFocus - not calculated and B_SHOT_PART_ON
restart the system.
If after restart not ok then call technical support
backward steps calculated in bGetUnitByStamp out of range restart the system
BBC_Acquisition Wnd - block creation failed
Action: restart the system
restart the system

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
BrandGE
ModelVoluson E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals