GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
3-20 Section 3-5 - Connection of Auxiliary Devices
6.) Press the (down) button repeatedly, until [MENU GRADE] is highlighted.
7.) Press the
(right) button and select [ENHANCED].
8.) Press the
(left) button, to exit the “Menu Grade” menu.
NOTE: Now additional items (see: Figure 3-15 on page 3-20) are displayed in the “Setup Menu”.
9.) Select [TV SYSTEM] by pressing the
(down) button.
10.)Press the
(right) button. The “TV System” Sub menu appears.
11.)Select the appropriate video signal by pressing the
(up) or (down) button respectively.
To check the video signal, see: Table 3-6, “TV-System and Region Code,” on page 3-21.
12.)Press the SET
button (C).
The message “Now loading...” appears and the system starts to save the settings.
13.)When “Power off” appears on the display, press the POWER
switch on the front of the DVD
recorder to turn power off.
14.)Press the POWER
switch again to turn power on.
Figure 3-14 “Setup Menu” - [MENU GRADE] highlighted
Figure 3-15 “Setup Menu” - [TV SYSTEM] highlighted
change MENU GRADE
to ENHANCED
Menu Grade: ENHANCED
TV SYSTEM
REGION CODE

Table of Contents

Other manuals for GE Voluson E8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson E8 and is the answer not in the manual?

GE Voluson E8 Specifications

General IconGeneral
BrandGE
ModelVoluson E8
CategoryMedical Equipment
LanguageEnglish

Related product manuals