3.6.2 The Receiving Analyzer
FIGURE 3-85
Confirmation or cancellation of the
transmitted setting information(1) is required
on the receiving Analyzer.
a) To accept data transfer of the category
indicated(1) from the sending Analyzer,
press Continue.
To cancel, press No.
b) If the transfer is accepted, while data is
being received, the following text will be
displayed:
Receiving via IR port…
with slight variations depending on the
category of settings information.
c) When the transfer is completed, the
following text will be displayed for a few
seconds:
Transmission OK
1 Settings to receive
Infra Red
Transmission
NoNo
ContinueContinue
Update
All Settings?
All current
settings will be
changed
1